MAURER - DIGITAL - 2-COLOR - PYROMETER
Series QKTRD 1085
Temperature range 600 to 3300°C (1112 - 5972°F)
Temperature control during production process
compact unit
Infrared - measuring transducer and electronic process unit in one case
|
- Vario-objective (vario-focus objective)
- light beam aiming device with green LED for target
- emissivity adjustable at the unit
- 2 - and 1 - color temperature
- analog- und digital output (serial interface)
- 1 limit output (open collector)
- software IR-LOG
|
Through the serial interface additional parameter functions are possible: |
analog output: 0 - 20 / 4 - 20 mA switchable |
zoom range within measuring range |
emissivity: 100,00 - 10,00 % |
average: arithmetical or sliding |
maximum value storage: storage modes and erase functions par ex. automatically with the next measuring object |
|
|
|
|
|
|
Examples for applications: |
steel, iron, non-ferrous metal, coating, wires, molding, hardening, induction heating, soldering, metal melt, forging, welding, transforming, vacuum furnace, rolling |
Technical datas: | |
Unit types |
QKTRD 1085-1 |
QKTRD 1085-2 |
Target marking |
light beam aiming device green LED or laser |
viewfinder |
Temp. measuring ranges: |
MR 1: 600 - 1600°C (1112 - 2912°F) MR 2: 750 - 2500°C (1382 - 4532°F) MR 3: 900 - 3000°C (1652 - 5432°F) MR 4: 1000 - 3300°C (1832 - 5972°F) |
Spectral range |
0,85 - 1,1μm and 0,95 - 1,1 μm |
Response time (t90) |
<1 ms with dynamical adaption |
Measuring uncertainity |
0,5 % of meas. value in °C ± 1°C (ε=1, Tu=23°C, T95 = 1s) |
Reproducibility |
0,1 % of meas. value in °C ± 1°C (ε=1, Tu=23°C, T95 = 1s) |
Emission ratio |
0,8 - 1,2 adjustable at the unit or through interface |
Emission factor ε |
100 - 10 % through interface |
Analog output |
0 - 20 mA oder 4 - 20 mA, load max. 500R |
Part measuring ranges: |
free adjustable within the measuring range |
Resolution |
<0,1% analog output, <0,1°C at interface |
1 limit output (open coll.) |
24 V DC / max. 100 mA |
Max. reading memory |
max.memory, erasing after time, external contact, by software, double memory |
Interface |
RS 232 isolated ± 50 V or RS 485 isolated 2500VRMS options: PROFIBUS, PROFINET, Ethernet, EtherCAT, USB 2.0 |
Software IR-LOG |
data recording, graph. representation, setting of pyrometer parameters |
adjustable parameter with Software IR-LOG |
emissionfactor,switching output,analog output, part meas.range, °C / °F, max.memory, average value, light beam aiming device switchable |
Objectives |
for accommodation to the measuring application an extensive selection of objectives are available |
Working temperature |
pyrometer 0 - 50°C (32 - 122°F), optic system up to 150°C (302°F) |
Stock temperature |
-10°C - +70°C (14 - 158°F) |
Temperature sensitivity |
0,05 % / °C |
Humidity tolerance |
35 - 85 % RF (non condensing) |
Operating Voltage |
24 V DC ± 10 % or 18 V AC ± 10 % < 160 mA |
Unit connection |
12-pole plug-connector |
Dimensions H / W / D |
54 x 54 x 147 mm (2,13 x 2,13 x 5,79 inch) / ALU-case |
Weight |
0,6 kg (1,32 lbs) |
Protection grade |
IP 65 |
Option |
built in digital display |
Datasheet
mechanical assembly | electrical assembly | electronic process unit |
execution in cooling case |
connection cable 12-pole |
AED 1012 |
Blowing device |
power supply 100-270 VAC - 24 VDC |
AED 1012-C PID controller
|
Mirror 90° |
digital display (built in-execution) |
AED 1012-PC program controller |
Mounting parts |
line scanner SC 1000 / 1012 |
|
|
PC-Box (USB – connection set) |
|
|